70
Preparing Your Laboratory
•
Pump Rate: 4 gal/min at 55 psi max
A PolyScience® WhisperCool™ Recirculating Chiller meets these requirements
and is recommended for the instrument. The PolyScience® WhisperCool™ is
available through PerkinElmer in the following two configurations: 208-230 V, 50
Hz (Part No. N0772045); 208-230 V, 60 Hz (Part No. N0772046).
An additional 200-240 V line is required for the chiller. A Hubbell® number 4560
receptacle (NEMA L6-15R configuration) is supplied with the PolyScience®
WhisperCool™ Chiller.
Handling of Gas Cylinders and Other Suggested Safety Practices
•
Fasten all gas cylinders securely to an immovable bulkhead or a permanent
wall.
•
When gas cylinders are stored in confined areas, such as a room, ventilation
should be adequate to prevent toxic or explosive accumulations. Move or store
gas cylinders only in a vertical position with the valve cap in place.
•
Locate gas cylinders away from heat or ignition sources, including heat lamps.
Cylinders have a pressure-relief device that will release the contents of the
cylinder, if the temperature exceeds 52 °C (125 °F).
•
When storing cylinders external to a building, the cylinders should be stored so
that they are protected against temperature extremes (including the direct rays
of the sun) and should be stored above ground on a suitable floor.
•
Mark gas cylinders clearly to identify the contents and status (full, empty, etc.).
•
Do not attempt to refill gas cylinders.
•
Arrange gas hoses where they will not be damaged or stepped on and where
things will not be dropped on them.
•
Perform periodic gas-leak tests by applying a soap solution to all joints and
seals.
•
Never view the ICP torch directly without protective eyewear. Potentially,
hazardous ultraviolet radiation may be emitted. Ordinary safety glasses will, in
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......