254
Maintenance
8. Replace the cover and return the assembly to the instrument.
9. Reattach the Axial Flat Optic Temperature Sensor to the instrument.
10. Return the thumbscrew on the heat sink cover.
11. Return the Axial Heat Sink Cooling Tube. See Remove and Replacing the
Axial Heat Sink Cooling Tube on page 247.
Nebulizers
PerkinElmer offers different types of nebulizers for a wide variety of applications.
For descriptions of the nebulizers and their different uses and advantages, refer to
Chapter 3, System Description. Part numbers and replacement parts for the
different types of nebulizers are listed on the next page.
Recommended flow rates for the nebulizers are given in the table below. If you find
that a flow rate is insufficient, this often indicates that the nebulizer has clogged
and requires cleaning. On GemTip and GemCone nebulizers, you should also
check that the argon gas fitting is properly tightened. If a space collar is used,
tighten the fitting with a wrench until you can no longer turn the knurled space
collar. If a space collar is not used, finger-tighten the fitting, then using a wrench,
turn clockwise 1/8-turn.
Table 5-5.
Nebulizer Flow Rates
Argon Flow Rate
Nebulizer
Typical Value
Operating Range
GemTip Cross-Flow
0.6 L/min
0.4 - 1.3 L/min
Low-Flow GemCone
0.6 L/min
0.4 - 0.95 L/min
High Flow GemCone
1.0 L/min
0.7 - 1.0 L/min
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......