164
Installation
Fitting and Connecting the Autosampler Sampling Probe
Caution
When you fit the sampling probe, take care
not
to press forcefully
in any direction on the autosampler arm since this can damage the
arm.
The sampling probe is a complete assembly. Do
not
attempt to
take the probe apart as this will damage it, and reassembly is not
possible.
Never
attempt to move the sampling probe holder along the
autosampler arm by hand. The autosampler may be damaged if the
tower is moved by hand when it is under system control.
Disconnect the power supply to the autosampler if you need to
manually move the tower
Attention
Lorsque vous correspondez à la sonde de prélèvement, prendre
soin de ne pas appuyer avec force dans une direction quelconque
sur le bras de l'échantillonneur automatique car cela peut
endommager le bras.
La sonde de prélèvement est un ensemble complet. Ne pas essayer
de prendre la sonde à part comme cela l'endommager et le
réassemblage est pas possible.
Ne jamais tenter
de déplacer le support de sonde de prélèvement
le long du bras du passeur à la main. L'échantillonneur
automatique peut être endommagée si la tour est déplacé à la main
quand il est sous le contrôle du système.
Couper l'alimentation à l'échantillonneur automatique si vous avez
besoin de déplacer manuellement la tour.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...