Troubleshooting
Plasma Ignition and Stability Problems
Plasma Ignition Problems
Possible reasons for unsuccessful ignition follow. Correct the problem, if
possible, and then try to ignite the plasma again.
•
If the plasma will not light check the
ignitor.
Open the door to the torch
compartment. Check that the ignitor contact finger is making contact with the
Problem
Possible Cause(s)
Corrective Action
Cannot get
suggested
detection
limits
(continued)
The standard may have too
high a concentration.
The standard should be 100 times the
estimated detection limit.
You may not be comparing
the appropriate value for the
wavelength you are using.
Confirm that you are using the appropriate
wavelength and recheck the value. Refer to
the Syngistix Software Guide on the
Syngistix Document Pack CD for details
..
General
performance
problems
Performance problems such
as poor precision or loss of
signal can often be traced to
the setup and condition of the
sample introduction system.
You are having difficulty
getting values for silicon in
samples with HF.
See the section
Introduction System on page 320
Use a sapphire injector.
Unexplained
background
shift
The torch or injector may be
coated with a deposit that
reflects light, resulting in a
periodic background shift.
Replace the torch and injector.
Table 6-16.
Performance Problems: Troubleshooting Table
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......