202
Maintenance
Purge Viewing Window/Tubes
Remove and examine the window or purge tube for deposits. If you notice a drop
in UV performance clean or replace the windows. Clean according to the
instructions given later in this chapter or replace as necessary.
Nebulizer
Check the nebulizer spray pattern with deionized water. Clean or replace the
nebulizer as necessary.
Spray Chamber
Check for leaks around the nebulizer and drain fitting. Inspect the spray chamber
for deposits and check the condition of the O-rings.
Peristaltic Pump
Check that the pump rollers are clean, not scored, free from deposits due to spills,
and move freely. Remove and clean the pump head if necessary. Replace the pump
rollers if necessary.
Periodically observe the drain tubing. Liquid should drain smoothly from the spray
chamber. If not, adjust the pump tension on the pump tubing.
Drains
Check the spray chamber drain fittings, tubing and connections and replace if
necessary.
General System Maintenance
•
Air filters:
clean or replace the spectrometer and generator air filters as
necessary.
•
Chiller:
Flush out the chiller every 6 months. Fill the chiller reservoir with the
coolant fluid that includes inhibitor (Part No. N0776200). The chiller reservoir
has a capacity of 4.2 liters (1.1 US gallons).
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......