205
Quick-Change Adjustable Torch Module
Quick-Change Adjustable Torch Module
The Quick-Change Adjustable Torch Module consists of the torch assembly, spray
chamber, and nebulizer. These components are easily removed from the system for
regular inspection and maintenance. Normal wear on these components is shown
by metal or salt deposits on the quartzware or injector, glazed or devitrified quartz,
worn O-rings, or clogged nebulizer tips.
All the O-rings in the sample introduction system must be in place and be in good
condition. Otherwise, pressure leaks may occur, which can cause difficulties in
igniting the plasma and poor performance.
It is good practice to have spare parts available before removing any of the parts of
this module, particularly the torch, since it may be accidentally broken.
Replacement parts are listed at the end of this chapter.
Spray Chamber
By handling the spray chamber oil and dirt can
accumulate on it and this will affect performance. To
assure proper wetting wipe the spray chamber with
acetone to remove all oils and dirt.
Pump Head
Remove and clean with water or a mild solvent.
Purge Viewing
Windows/Tubes
Clean with deionized water and dry with a soft cloth.
To clean difficult stains, refer to the information in this
chapter.
Table 5-1.
Recommended Cleaning Procedure
(Continued)
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......