220
Maintenance
*See the note below.
2. Slide in the injector assembly straight into the bottom of the torch cassette then
rotate to the right to lock in place.
3. Tighten the knurled nut.
Replacing the Torch on the Mount
The following steps describe how to replace the torch in the mount. Refer to the
next two figures.
3
Knurled Nut (Part No. N0776027)
8
Cassette (Part No. N0810214)
4
Washer (Part No. N0776028)
9
Stripe on Torch
5*
O-Ring (Part No. 09902223) After
removing the torch the lower O-ring
may move out of position. Make
sure to push the lower O-ring all the
way down on the ledge it rests on
before you return the torch
glassware. If the O-ring is not in the
proper location leaks will occur.
10
Index Mark on Torch Holder
Note
If the torch glass does break, discard the broken torch and replace the torch
glass. Remove the knurled nut, sleeve and washers and make sure that you
remove all glass fragments
.
Caution
Attention
Only
use PerkinElmer Flat Plate Torch Glassware
TM
.
Using any other
glassware will damage the torch.
Utiliser uniquement PerkinElmer Plate Flat Torch Glassware
TM
.
L'utilisation de tout autre verrerie endommager la torche.
Item
Description
Item
Description
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......