18 Calibrating universal inputs
210
•
free chlorine, ph/Temp.-compensated
In the universal inputs calibration default settings for the "free chlorine, ph/T-
compensated" operation mode, the slope calibration is enabled and pre-config-
ured as the only available calibration routine.
Additional calibration default settings for calibration of chlorine sensors are ex-
plained in the following table.
Sample screen:
Universal input calibration
default settings in the
"free chlorine, pH/T-com-
pensated" operation
mode
Parameter
Possible settings
Explanation
Temperature com-
pensation
Selection from Analog selection
Temperature input for automatic sens-
ing of the test/sample solution tempera-
ture during the calibration
pH compensation
source
Selection from Analog selection
pH measuring input for automatic sens-
ing of the test/sample solution pH value
during the calibration
NOTE!
The "Calibration Default Settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
v
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......