7 Startup
62
7.2
Function test
Important hardware information is displayed in the "Device Info" menu.
The functionality of the device can be checked from here.
7.2.1
Checking the optional boards
Open the hardware information for the installed optional boards as follows:
Device Menu
r
Device Info
r
Slots
A tab with hardware and software information for each correctly installed option-
al board now appears.
If no tab appears for an optional board, the board was not recognized and there
is a hardware problem. In such cases, check whether the optional board con-
cerned was installed correctly.
v
Chapter 11 "Retrofitting optional boards", Page 161
If this is not successful, contact Technical Support at JUMO.
The contact data can be found on the back of these installation instructions.
7.2.2
Checking sensors and inputs/outputs
To check correct operation of all inputs/outputs, you can display the current an-
alog and binary values.
Device Menu
r
Device Info
r
Inputs/Outputs
Depending on the type of input, up to 2 columns will be visible in the Device Info
screens for inputs.
•
Compensated:
Display value that is calculated from the value measured by
the sensor through use of an appropriate compensation method and after
taking the corresponding calibration values into account.
This suppresses incorrect measuring values that can be caused by outside
factors (e.g. temperature) or sensor wear (e.g. a dirty electrode).
•
Uncompensated:
Value measured by sensor (raw value of the measure-
ment input, e.g. pH measuring chain voltage)
These sensor values are subject to distortion caused by outside factors.
Sample screen:
Hardware Information
for an optional board
"Universal input"
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......