17 Calibrating Ci conductivity sensors
17.3.2
Calibrating the temperature coefficient
Step
Action
1
Start calibration of the temperature coefficient.
Device Settings Menu
r
Calibration
r
Select Ci Analysis Input or
Universal Input
r
TC Calibration
2
Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
for Ci optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" operat-
ing mode, continue with step 4
3
Enter one of the measuring ranges 1 to 4. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
tap button to enter
measuring range
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......