6 Electrical connection
IN 7
IN 8
pH/Redox
Symmetric connection of a combination electrode with
integrated resistance thermometer and VarioPin plug
head
Symmetric connection is used to reduce interference from
stray electromagnetic fields along the sensor cable.
The RTD temperature probe is used to provide a tempera-
ture-compensated pH-value measurement, and can be con-
nected to a temperature input or universal input.
A = Glass/metal electrode (core)
B = Reference electrode (inner shield)
C = Liquid potential (grounding pin, pipe, or container wall at
the measuring point)
D = Shield (outer shield)
E = RTD temperature probe
F = RTD temperature probe
G = RTD temperature probe
Terminal 2 is not connected!
Slot
Option/connection variant
Symbol
?
A
B
E
F
G
D
C
1
3
5
6
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......