10 Configuration
124
Connection type
2-wire,
3-wire
only for signal types Pt100, Pt1000, 400
Ω
and
4000
Ω
:
Connection variants for the connected resistance
thermometer
Customer-specific
linearization
Selection of a
linearization table
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To cre-
ate a linearization table, you need the JUMO PC
setup program.
Unit
up to 5 text characters
Process variable unit
not adjustable for pH measurement
The temperature unit is specified in the basic set-
tings.
v
Chapter 10.2 "Basic settings", Page 116
Scale start
-99999 to +99999
b
only for standard signals:
Measured value from sensor (uncomp.) that cor-
responds to the lower limit of the standard signal
range [0 V or 0(4) mA];
Refer to the technical data for
the sensor.
Scale end
-99999 to +99999
b
only for standard signals:
only for standard signals:
Measured value from sensor (uncomp.) that cor-
responds to the upper limit of the standard signal
range [10 V or 20 mA]
Refer to the technical data for
the sensor.
Start of
display range
-99999 to +99999
b
Upper/lower limit for labeling the scale when dis-
playing measuring values, e.g. in recorder dia-
grams and bar graphs
Display range end
-99999 to +99999
b
Decimal place
Auto
Fixed comma format
Decimal points in the display
Offset
-999 to +999
b
only for temperature and
conductivity measurement:
Correction value added to measuring value
Filter time constant
0.0 to 25.0 s
Optimization of measuring value updating
The higher the value of the filter time const., the
slower the measured value is displayed.
Ra
0 to 4000
Ω
only for IN 11/12:
Resistance value between the
slider (S) and start point (A) in a resistance ther-
mometer/resistance potentiometer when the slid-
er is at the start point
Configuration point Selection/
setting option
Explanation
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......