11 Retrofitting optional boards
164
11.2
Ci base calibration
Analysis inputs for inductive conductivity sensors must undergo a Ci base cali-
bration in the course of commissioning. A basic Ci calibration must be performed
when:
•
A new sensor or a new Ci optional board is being installed for the first time
•
The sensor or a Ci optional board was replaced
•
A Ci optional board was repositioned to a different optional board slot
•
Data loss occurred as the result of failure of the backup battery while the
power was switched off
•
Device software was updated
After the basic calibration has been performed, the measuring input can be cal-
ibrated. Following successful calibration, the measuring input is ready for use.
NOTE!
For the Ci base calibration, you need the JUMO Type 202711/21 calibration
adapter for inductive conductivity sensors (TN 00543395).
CAUTION!
A buffer battery is built into the device. Its purpose is to maintain data when the
device is switched off or if the voltage supply is interrupted. If the battery ap-
proaches the end of its operating life (approximately 7 years) then a battery
prealarm will indicate that status. When the battery is empty a battery alarm will
indicate that status. The battery must be exchanged before it is empty. The bat-
tery must be exchanged by the JUMO Service department! In this case, send
in the device!
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......