16 Calibrating CR conductivity sensors
190
16.2.2
Calibration default settings for CR conductivity sensors
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the CR calibration are described
in the following.
Open the calibration default settings:
Device menu
r
Calibration
r
Select CR Analysis Input
r
Calibration Default Settings
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
NOTE!
The "Calibration Default Settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
v
Chapter 8.2.1 "Log-on/Log-out", Page 78
Sample screen:
CR calibration default
settings
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......