141
10 Configuration
Man. output level 2
0 to 100%
only for coarse/fine controllers:
Preconfigured Man. output level for the 2nd con-
troller output
Accepted automatically upon activation of the
manual mode
mode if acceptance of the manual output level
is set to "Yes"; otherwise the controller
assumes the last output level from the automatic
control mode.
Acceptance of hold
output level 1
Yes
No
Acceptance of the preconfigured hold output (for
the 1st controller output on coarse/fine control-
lers) upon activation of the hold mode
Hold output level 1
-100 to +100%
Preconfigured hold output level (for the 1st con-
troller output on coarse-/fine controllers)
Accepted automatically on activation of the
"manual mode" if acceptance of the manual out-
put level is set to "Yes".
Otherwise, the controller accepts the last output
level from the automatic control mode.
Acceptance of hold
output level 2
Yes
No
only for coarse/fine controllers:
Acceptance of the preconfigured hold output for
the 2nd controller output upon activation
of the hold mode
Hold output level 2
0 to 100%
only for coarse/fine controllers:
Preconfigured hold output level for the 2nd con-
troller output
Accepted automatically upon activation of the
hold mode if acceptance of the hold output level
is set to "Yes"; otherwise, the controller assumes
the last output level from the automatic control
mode.
Self-optimization
Enabled
Disabled
Enabling of self-optimization
Alarm monitoring
Inactive
Active
Activation/deactivation of alarm monitoring
Alarm monitoring provides a continuous plausi-
bility check of the control deviation. A dosing
alarm is triggered when the control deviation is
greater than the value of the alarm tolerance set
in the controller parameters. If "Alarm ack." is ac-
tivated (see next configuration item), the control-
ler switches to hold mode on a holding alarm.
Configuration point Selection/
setting option
Explanation
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......