6 Electrical connection
36
Use shrink tubing to insulate the braided shield
Preparing a shielded multiconductor cable
Strip the insulation off the connecting cable and pull back the shield
Attach wire ferrules to the ends of the conductors and a Shield-Kon® connector
1
for shielding
Inner conductor
Shrink sleeve
NOTE!
The black, semiconducting layer must not touch with the inside conductor! This
would short-circuit the signal from the pH electrode.
1.Shield-Kon is a registered trademark of THOMAS & BETTS INTERNATIONAL, Inc.,
Wilmington Del., US.
60 mm
60 mm
Shield-Kon® connector