15 Calibrating ammonia sensors
15.1
Notes
15.2
General information
The calibration of ammonia sensors is based on measurements in ammonia-
free test solutions.
15.2.1
Calibration methods for ammonia sensors
Zero-point calibration
This calibration method is used to determine the ammonia zero point.
An ammonia-free test solution (e. g. water) is needed as a reference.
15.2.2
Calibration default settings for ammonia sensors
In the ammonia sensor calibration default settings, the zero-point calibration is
enabled and preconfigured as the only available calibration routine.
Open the calibration default settings:
Device menu
r
Calibration
r
Select Analysis Input for pH/Redox/
NH
3
r
Calibration Default Settings
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
Sample screen:
Ammonia calibration
default settings
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......