16 Calibrating CR conductivity sensors
Calibration default settings for calibrating the relative cell constant
Calibration default settings for calibrating the temperature coefficient
Parameter
Setting options
Explanation
Reference conductivi-
ty
0 to 9999 mS/cm
Conductivity of the reference solution
Parameter
Setting options
Explanation
Temperature com-
pensation
Selection from
analog selection
Temperature input for automatic sensing of the test/
sample solution temperature during the calibration
Reference tempera-
ture
-50 to +150° C
The conductivities of a sample solution at the refer-
ence temperature and operation temperature are
sensed during the calibration procedure. This yields
2 value pairs (temperature/conductivity). These val-
ue pairs provide the basis for calculating the tem-
perature coefficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
Operation tempera-
ture
-50 to +150° C
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......