17 Calibrating Ci conductivity sensors
17.1
Important information
17.2
General information
The calibration of Ci sensors is based on measurements in test solutions with a
defined electrolytic conductivity. Since the electrolytic conductivity of liquids is
temperature dependent, the temperature of the test solution must be sensed.
This requires either manual entry or measurement with the aid of a temperature
sensor.
17.2.1
Calibration methods for Ci conductivity sensors (inductive)
Rel. cell constant
The deviation from the nominal cell constant of a Ci sensor is described by the
relative cell constant. The rel. cell constant is determined by making a measure-
ment in a test solution with a defined conductivity.
Temperature coefficient
The temperature coefficient is a measure of the temperature dependence of the
electrolytic conductivity of a liquid. Is used to compensate for the effect of tem-
perature when measuring the electrolytic conductivity. When making a tempera-
ture-compensated conductivity measurement, the measuring value of the
conductivity is always given on the basis of the fixed, preset reference tempera-
ture. With the aid of the temperature coefficient, the value of the electrical con-
ductivity at the reference temp. is calculated from the current values of
conductivity and temperature measured for the liquid and then displayed.
Reference temperature is set in the configuration of the individual Ci analysis in-
put.
v
Chapter 10.5.4 "CR/Ci analysis inputs (conductive/inductive conductivity)",
Page 128
The temperature coefficient is determined from 2 measurements in a test solu-
tion at different temperatures (reference and operation temp.).
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
NOTE!
If the temperature coefficient of a sample solution is known, it can be entered
directly.
v
Chapter 12.2.1 "General procedure for calibration", Page 169
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......