6 Electrical connection
38
6.2.3
Conductor cross-sections for optional boards
The terminals on optional boards are plug-in screw terminals.
Optional boards for
Ferrule
Conductor cross sec-
tion
Length to
strip
minimal
maximal
Universal inputs
Analog outputs
Binary inputs
Digital outputs PhotoMOS®
a
Logic outputs
Voltage supply output
without ferrule
0.14 mm
2
1.5 mm
2
7 mm
with ferrule with lip 0.25 mm
2
0.5 mm
2
7 mm
Ferrule without lip
0.25 mm
2
1.5 mm
2
7 mm
Rigid
0.14 mm
2
1.5 mm
2
7 mm
Analysis inputs pH/Redox/NH
3
Analysis inputs CR
b
Analysis inputs Ci
c
Digital outputs relay
Digital outputs triac
without ferrule
0.2 mm
²
2.5 mm²
7 mm
with ferrule with lip 0.25 mm
2
1.5 mm
2
7 mm
Ferrule without lip
0.25 mm
2
2.5 mm²
7 mm
Rigid
0.2 mm
²
2.5 mm²
7 mm
a
PhotoMOS®
is a registered trademark of Panasonic.
b
CR analysis inputs = Analysis inputs for resistive conductivity
c
Analysis inputs Ci = Analysis inputs for inductive conductivity
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......