11 Retrofitting optional boards
6
Insert the optional board in the selected slot. Ensure that the board
is seated correctly.
7
Fill all empty slots with plastic board frames.
8
Re-insert the rear panel of the case and fasten with 3 screws. Make
sure that all the toothed lock washers for securing the screws are
re-inserted (see Step 7). The screws represent the protection con-
ductor connection to the rear panel of the case.
9
Reinsert all the plug-in screw terminals and interface cables that
were removed in Step 4 into the correct sockets.
For "Ci analysis input" optional boards (inductive conductivity),
continue at Step 10; otherwise, at Step 12
10
Connect the inductive conductivity sensor using the M12 socket
adapter from the scope of delivery of the Ci optional board to the
Ci optional board of the device.
11
Connect the 2-wire temperature sensor cable of the M12 socket
adapter to an appropriate analog input (e. g. temperature measur-
ing input). Observe the specifications for the temperature sensor
integrated into the conductivity sensor.
15
Now, reconnect the power supply and check whether the new
hardware has been recognized.
v
Chapter 7.2.1 "Checking the optional boards", Page 62
16
Only for Ci optional boards (inductive conductivity):
Perform a Ci base calibration.
v
Chapter "Performing the Ci base calibration", Page 165
17
Only for optional boards for analysis inputs:
Calibrate the analysis inputs.
v
Chapter 12 "Calibration in general", Page 169
Step
Action
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......