298
Analog inputs
Proper operation of the analog inputs of the device can be checked here.
In addition to selecting the analog input and signal type, the line frequency of the
power supplied to the device must be set in order to test an analog input correct-
ly. Pressing the "Test" button displays a connection diagram for the test signal.
Ensure that a valid test signal acc. to the connection diagram is present at the
input. Confirming the connection diagram starts the test. The measuring values
coming from the analog input concerned can be viewed in the actual-value dis-
play and in the measuring value table. To stop the test, press the "Stop" button.
Open:
Online Parameters
r
Adjusting / Testing (double-click)
r
Analog inputs tab
(1)
"Stop" button
Stops a test
(2)
"Test" button
Starts a test
(3)
Actual-value display
Displays the actual analog value at the input
(4)
Measuring value table
Lists the measuring values over the course of time
(5)
"Supply frequency" setting
(6)
Selection of the analog input to test
(7)
Selection of the signal type to test
(1)
(7)
(6)
(5)
(2)
(3)
(4)
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......