8 Operation
Acknowledging collective/dosing alarms
The
collective alarm
combines all alarms in the alarm list. It simplifies signaling
of one or several active alarms with external indicating devices or to control
rooms. The digital signal for the collective alarm is available in the binary selec-
tor in 2 versions:
•
Collective alarm:
Is triggered by the occurrence of any alarm and disap-
pears only after all alarms in the alarm list have been cleared.
•
Collective alarm ack.:
Is triggered by the occurrence of any alarm and dis-
appears when the alarm is acknowledged.
Selection of a collective alarm when configuring device functions and
digital outputs:
Binary selection
r
Alarm and internal -signals
r
Collective alarm/Collective alarm ack.
Dosing alarms
are triggered by the controllers if the absolute value of the con-
trol deviation is larger than the "alarm tolerance". An "alarm delay" can be set for
each dosing alarm. Dosing alarms and acknowledgment of them must be acti-
vated in the controller configuration. The "alarm tolerance" and "alarm delay"
settings are made in the controller parameters.
Chapter 10.11.1 "Configuration of the controllers", Page 140
Chapter 9.2 "Parameter Sets (Controller parameter)", Page 110
Selection of a dosing alarm when configuring internal functions and digital out-
puts:
Binary selector
r
Controller
r
Dosing Alarm Controller 1 to 4
Collective and dosing alarms can be acknowledged with the "Confirm" button.
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......