14 Calibrating Redox sensors
14.3
Redox calibration routines
14.3.1
Zero-point calibration
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
v
Chapter 8.2.1 "Log-on/Log-out", Page 78
Step
Action
1
Ensure that
•
the calibration default settings are correct,
•
"mV" is set as the Redox unit in the configuration of the Redox
measuring input.
v
Chapter 14.2.2 "Calibration default settings for Redox sensors",
v
Chapter 10.5.3 "pH/Redox/NH3 analysis inputs", Page 126
2
Start the Zero-point calibration.
Device Menu
r
Calibration
r
Select Analysis Input for pH/Redox/
NH
3
r
Zero-Point Calibration
3
Check that the "Redox test solution" value displayed matches the
Redox value of the test solution.
The "Redox test solution" value is taken from the calibration default
values. It is still possible to change this manually here.
4
Clean the Redox electrode and immerse it in the test solution. Wait
until the measuring value displayed stabilizes and then confirm the
result of the measurement by pressing "OK".
5
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6
Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.
Tap button to
change the Redox
value of the test
solution manually
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......