125
10 Configuration
Rs
6 to 4000
Ω
only for IN 11/12:
Span of the variable
resistance value between the slider (S) and
start point (A)
Re
0 to 4000
Ω
only for IN 11/12:
Resistance value between the
slider (S) and end point (E) in a resistance ther-
mometer/resistance potentiometer when the slid-
er is at the end point.
Compensation tem-
perature
Selection from
analog selection
Analog input of the compensation thermometer
for temperature-compensated measurement of
pH value, free chlorine or conductivity
Compensation
TC linear,
TC-curve,
natural water,
natural water with
expanded temperature
range,
ASTM neutral,
ASTM acid,
ASTM alkaline,
NaOH 0 to 12%,
NaOH 25 to 50%,
HNO
3
0 to 25%,
HNO
3
36 to 82%,
H
2
SO
4
0 to 28%,
H
2
SO
4
36 to 85%,
H
2
SO
4
92 to 99%,
HCL 0 to 18%,
HCL 22 to 44%
Type of temperature compensation for conduc-
tivity measurement
Reference temp.
15 to 30 °C
required only for conductivity measurement
with "TC linear" or "TC curve" temperature
compensation:
Temperature at which the (temperature-compen-
sated) conductivity value displayed was set
Compensation
pH value
Selection from
analog selection
Analog input of the pH-value sensor for pH-com-
pensated measurement of free chlorine
Alarms 1/2
Analog input alarms are used to monitor measured values
in relation to adjustable limit values.
The alarm settings for all analog device functions are explained together.
v
Chapter 10.9.2 "Alarms for analog signals", Page 135
Calibration timers
Calibration timers prompt the user to calibrate sensors on a regular basis.
The settings for all analysis inputs and the universal inputs are explained to-
gether.
v
Chapter 10.10 "Calibration timers", Page 139
a
Resistance potentiometer
b
The unit for the particular sensor value appears in the entry field.
Configuration point Selection/
setting option
Explanation
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......