10 Configuration
138
10.9.3
Digital signal alarms
Open:
Device menu
r
Configuration
r
Digital inputs
r
Digital input 1 to 9
Configuration point Selection/
setting option
Explanation
Digital signal for hold Selection from the
binary selector
Digital signal for activating the hold function
When the hold function is activated, the alarm as-
sumes the state defined in the "Response at
hold" setting.
Response at hold
inactive
active
hold
normal
Specification of the alarm state when the hold
function is activated
inactive:
alarm suppressed
active:
alarm forced
hold:
alarm state is maintained regardless of
alarm condition changes
normal:
alarm in acc. with alarm condition
Alarm type
active
inactive
Arming or disarming an alarm
Alarm/Event list
off
event
alarm
Assignment of alarm notification to alarm list or
event list
Alarm text
up to 21 text characters
Text for the alarm/event list
Polarity for alarm
high
low
Digital input alarm condition
Alarm delay
0 to 999 s
Time delay between occurrence of the alarm
condition and triggering of the alarm
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......