165
11 Retrofitting optional boards
Performing the Ci base calibration
Step
Action
1
Ensure that you have user rights for calibration settings.
The factory settings authorize the "Master" and "Service" users for
this.
v
Chapter 8.1.1 "Passwords and user rights", Page 65
2
Ensure that the electronics of the JUMO AQUIS touch P have
reached operating temperature. You can view the board tempera-
ture at:
Device Menu
r
Service
r
Service Data
r
"Internal Data" tab
Ensure that the ambient temperature specified for the device cor-
responds to the conditions for the normal operating mode. Wait un-
til the board temperature has reached an approximately constant
value.
3
Place the sensor such that the sensor body is suspended freely in
air. Observe the following rules during the entire calibration:
•
Keep all objects away from the sensor body
•
Do not touch the sensor body
•
Do not allow the sensor body to lie flat on a surface
Sensor body of a Ci
sensor
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......