5 Mounting
28
5.2
Dimensions
Spacings
For adequate ventilation, the following mounting distances from adjacent equip-
ment must be respected:
•
horizontal 35 mm
•
vertical 80 mm
When inserting the device, sufficient space for maneuvering the connecting ca-
bles must be left behind the case. A Ci-conductivity sensor is connected via a
M12 connector adapter. This increases the space requirement with respect to
mounting depth (see dimensional drawing below).
92 mm
+0.8 mm
92 mm
+0.8 mm
130,9 mm
96 mm
123,1 mm
90,5 mm
97,1 mm
104,1 mm
96 mm
22,25 mm
22,25 mm
22,25 mm
22,25 mm
90,5 mm
97,1 mm
104,1 mm
5,5 mm
Panel thickness max. 5 mm