173
12 Calibration in general
Assessment criteria
pH calibrations
(glass electrodes and ISFET connected to analysis measuring inputs as well as standard sig-
nals connected to universal inputs)
pH calibrations (antimony electrodes connected to analysis measuring inputs)
Redox zero-point calibration
Ammonia calibration
Calibration value [unit]
—
—
Zero point [pH]
...
<
5
≤
...
<
6 to 8
<
...
≤
9
<
...
Slope [%]
...
<
75
≤
...
<
89.6 to 103.1
<
...
≤
110
<
...
Calibration value [unit]
—
—
Zero point [pH]
...
<
-2 to +2
<
...
Slope [%]
...
<
10 to 110
<
...
Calibration value [unit]
—
—
Zero point [mV]
...
<
-200
≤
...
<
-120 to +120
<
...
≤
+200
<
...
NOTE!
There is no assessment of the calibration values in the case of a Redox 2-point
calibration.
Calibration value [unit]
—
—
Zero point [mV]
...
<
-612
≤
...
<
-312 to +588
<
...
≤
+888
<
...
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......