![JUMO AQUIS touch P 202580 Скачать руководство пользователя страница 184](http://html1.mh-extra.com/html/jumo/aquis-touch-p-202580/aquis-touch-p-202580_operating-manual_2031168184.webp)
14 Calibrating Redox sensors
184
14.2.2
Calibration default settings for Redox sensors
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the Redox calibration are de-
scribed in the following.
Open the calibration default settings:
Device Menu
r
Calibration
r
Select Analysis Input for pH/Redox/
NH
3
r
Calibration Default Settings
Redox calibration default settings
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
NOTE!
The "Calibration Default Settings" menu appears in the Device Menu only if a
user with corresponding user rights is logged in.
v
Chapter 8.2.1 "Log-on/Log-out", Page 78
Sample screen:
Redox calibration default
settings
(zero-point calibration)
Parameter
Possible settings
Explanation
Redox test solution
-1500 to +1500 mV
Manual entry of the Redox potential of
the test solution being used for calibra-
tion
NOTE!
Remember that the configuration of the Redox measuring input must be set to
the Redox unit "mV" for the zero-point calibration and to "Percent" for the two-
point calibration.
v
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......