16 Calibrating CR conductivity sensors
192
16.3
CR calibration routines
16.3.1
Calibrating the relative cell constant
NOTE!
Conductivity measuring inputs can be configured with measuring range change-
over. Accordingly, calibrations must be performed for all "accessible measuring
ranges".
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
v
Chapter 8.2.1 "Log-on/Log-out", Page 78
Step
Action
1
Start calibration of the relative cell constant.
Device menu
r
Calibration
r
Select CR Analysis Input
r
Rel. cell
constant Calibration
For CR optional boards, continue with step 2; for universal inputs
set to the "conductivity measurement" operation mode, continue
with step 3
2
Enter one of the measuring ranges 1 to 4. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
Tap button to enter
measuring range
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......