309
20 Technical data
20.2.6
Measuring circuit monitoring, optional boards
Input/Sensor
Underrange/
overrange
Short circuit/
sensor break
Open circuit
Special fea-
tures
pH-value (glass elec-
trode)
Yes
Configurable
Impedance mea-
surement
a
a
With impedance monitoring and detection of deposits, the sensor alarm is tripped in case of a fault.
Monitoring via impedance measurement can also be activated.
The following points should be observed to ensure correct function:
Impedance measurements are possible only with glass-based sensors.
Sensors must be connected directly to an analysis input for pH/Redox/NH
3
on the device.
Impedance converters must not be installed in the measuring circuit.
The maximal admissible cable length between sensor and device is 10 m.
Liquid resistance has a direct impact on the measurement result. It is therefore advisable to activate the im-
pedance measurement in liquids at a minimum conductivity of approx. 100 µS/cm.
Configurable
Impedance mea-
surement
a
-
pH-value (ISFET)
Yes
No
b
b
Errors in the measuring circuit (short circuit or line break) lead to display errors (underrange or overrange
or inadmissible value).
No
b
-
Resistive conductivity
Yes
No
b
Configurable
Only with
4-wire circuit
a
Inductive conductivity
Yes
No
b
No
b
-
Universal input for con-
nection of: voltage/cur-
rent signal, RTD
temperature probe
Yes
No
b
No
b
-
Universal input for con-
nection of: resistance
potentiometer
No
b
No
b
No
b
-
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......