10 Configuration
10.7
Digital inputs of base unit and optional boards
Open:
Device menu
r
Configuration
r
Digital inputs
r
Digital input 1 to 9
Response at hold
low
high
NAMUR low
NAMUR high
hold
safety value
Specification of the analog output value when the
hold function is activated, during calibration of
one of the sensors for the particular output or at
error (overrange/underrange)
low:
lower limit of the standard signal value
range
[
0 V or 0(4) mA
]
high:
upper limit of the standard signal value
range (10V or 20 mA)
NAMUR low:
lower NAMUR limit of the standard
signal
[
0 V or 0(3.4) mA
]
NAMUR high:
upper NAMUR limit of the stan-
dard signal (10.7 V or 22 mA)
hold:
unchanging analog value
Safety value:
see configuration item
"Safety value" in this table
Response at calibra-
tion
Moving
Frozen
Safe value
Response at error
low
high
NAMUR low
NAMUR high
hold
safety value
a
The unit for the value set for the "Signal output value" appears in the entry
field.
Configuration point Selection/
setting option
Explanation
Configuration point Selection/
setting option
Explanation
Description
up to 21 text characters
Designation for the input
Inversion
Yes
No
Invert/do not invert state
Contact
Base unit:
potential-free contact,
external voltage source
Type of digital signal connected
Optional boards:
potential-free contact
Alarm
Digital inputs alarms are used to monitor input-side switching signals.
The alarm settings for all digital device functions are explained together.
v
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......