83
8 Operation
8.2.5
Service
The "Service" menu is used for in-device troubleshooting and diagnostics. It is
intended primarily for trained personnel. When troubleshooting together with
Service, the operator can access data that the JUMO service technicians need
for diagnosis.
Service Menu items
•
Default-configuration:
The operator can store the current device configu-
ration in the device's memory. This configuration can be reloaded as the ac-
tive configuration at any time. This is important, for instance, to return to the
initial configuration after configuration changes made for testing purposes.
Configurations can also be saved with the PC setup program or on a USB
flash drive with the aid of the Memory Manager.
v
Chapter 8.4 "Memory Manager (USB flash drive)", Page 90
•
Debug Window:
The Debug window can be used to check the behavior of
the device software in detail. Data are displayed to assist the JUMO service
technicians during troubleshooting.
•
Ci base calibration:
The Ci base calibration is required when commission-
ing an optional board or sensor for inductive conductivity measurements and
can be performed here.
v
Chapter 11.2 "Ci base calibration", Page 164
•
Service data:
Status information
that can be evaluated by JUMO service
personnel for diagnostic purposes can be read here.
In addition, the
"Service count."
and
"Internal data"
are displayed. The
NOTE!
The Service menu appears in the Device menu only if a
user with corresponding user rights is logged in.
v
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......