269
19 PC Setup Program
19.8.9
Buffer set tables
Three Buffer set tables are created in the JUMO AQUIS touch P.
Buffer set tables are needed for calibration of pH sensors when using automatic
buffer detection.
v
Chapter 13 "Calibrating a pH measuring chain", Page 175
.
These Buffer set tables can be changed as often as necessary here.
The factory settings for the Buffer set tables are as follows:
•
Buffer set 1:
Reference buffer solutions for calibrating pH measuring instru-
ments acc. to DIN 19266
•
Buffer set 2:
Technical buffer solutions, preferably for calibrating and adjust-
ing technical pH measuring instruments acc. to DIN 19267
•
Buffer set 3:
No data
Open Settings:
Only Setup
r
Buffer set tables
r
"Buffer set table 1 to 3" tab
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......