10 Configuration
142
Pulse length output
With a pulse width output, the output level generated by the controller has the
form of the pulse/pause ratio of a square-wave signal with a fixed frequency
(pulse width modulation). Output of this type of signal from the device uses a dig-
ital output.
Alarm ack.
Inactive
Active
Activation/deactivation of the acknowledgment
function
Dosing alarms of the respective controller must
be acknowledged in the "Alarm list" when "Alarm
acknowledgement" is activated. Dosing alarms
are not cleared automatically when the control
deviation drops to a value that is less than or
equal to the alarm tolerance. The alarm tolerance
is set in the controller parameters.
v
Chapter 9.2 "Parameter Sets (Controller pa-
Configuration point Selection/
setting option
Explanation
100%
50%
0%
Cy
10%
90%
90%
10%
T
ON
50%
50%
T
Off
10%
90%
0
1
w
X
P
X
P
x
-
w
x
Y
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......