6 Electrical connection
50
IN 7
IN 8
Ci optional board (inductive conductivity measurement)
Connection via an M12 plug,
Attach connections for compensation
thermometer (2-wire cable for socket) to a
suitable analog input (2-wire circuit),
The factory-installed wiring must not be changed!
CR optional board (resistive conductivity measurement)
2-electrode system with 2-wire conductor
Terminal 1 must be connected to the
outer electrode on concentric conductivity sensors.
A = Outer electrode (core color for JUMO types with fixed ca-
ble: White)
B = Inner electrode (core color for JUMO types with fixed ca-
ble: Brown)
C = Shield
CR optional board (resistive conductivity measurement)
2-electrode system with 4-wire conductor
(Wiring to minimize the measuring error caused by lead-wire
resistance)
Terminal 1 must be connected to the
outer electrode on concentric conductivity sensors.
A/B = Outer electrode
C/D = Inner electrode
E
= Shield
Slot
Option/connection variant
Symbol
C
B
A
1
2
3
4
6
E
C
D
B
A
1
2
3
4
6
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......