Section 8-2 - External Component Replacement Procedures
8-2-4-5
Control Panel and Keyboard Installation Procedure
NOTE:
When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the
Control Panel and Keyboard Removal
NOTE:
When installing the Control Panel and Keyboard during other replacement procedures, Step 1
will only be required if indicated.
1) Stand the replacement Control Panel and Keyboard beside the Vivid q N system and tilt it on its
side to gain access to the cable connector underneath. Re-connect the keyboard/control panel
cable from the system to the cable connector (refer to
2.) Carefully lift the Control Panel and Keyboard into position on the Vivid q N (refer to
above). Make sure it is properly seated and take care to align the screw holes with those in the
Bottom Assembly.
3.) Close the Vivid q N LCD Display cover, making sure both front latches are securely fastened.
4.) Carefully turn the Vivid q Nover, placing it face down.
5.) Return and fasten each of the 9 screws that secure the Control Panel and Keyboard onto the
Bottom Assembly (refer to
6.) Re-install the battery as described in the
"Battery Installation Procedure" on page 8-5
.
7.) Return the Handle as described in the
"Bearing Handle Installation Procedure" on page 8-3
.
8.) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
9.) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
- only if the Vivid q N system is
currently mounted on the SafeLock cart
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
-
"Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
-
"Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
-
"Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
-
"Alphanumeric Keyboard Test" on page 4-19
-
"External Keyboard Test" on page 4-20
-
-
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...