Chapter 3 - System Setup
3-115
c.) Edit the name and/or the IP address of the server, see
d.) Click
OK
to save the new settings and close the Edit Name dialog box.
e.) Click
OK
to close the IPs dialog to return to the Properties dialog box.
f.) Enter the DICOM server AE Title. This entry is case-sensitive and must match exactly.
g.) Enter the DICOM server port.
h.) For some DICOM Servers, the default Time-out setting (30) is too low. Change if necessary
by typing a new value (seconds).
i.)
Click
OK
to close the Worklist Properties dialog box and save the changes.
Figure 3-123 Edit Name and/or IP address
Figure 3-124 DICOM Server’s Properties Dialog Box
Name
IP address
AE Title
Port No
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...