Section 4-2 - General Procedures
3.) The software initiates and sets up the Front-End electronics and the rest of the scanner.
4.) The Keyboard backlight is lit.
5.) As soon as the software has been loaded, a 2D screen is displayed on the monitor, as
shown below.
NOTE:Total time used for start-up is typically three minutes.
Figure 4-3 Start-up Progress Bar for Vivid
i
n
Figure 4-4 2D Screen