8-60
Section 8-3 - Internal Component Replacement Procedures
8-3-9-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the AUO LCD Flex Cable replacement
procedure.
8-3-9-2
Time Required
30 minutes
8-3-9-3
Preparation
Shut down the Vivid q N ultrasound unit, as described in
8-3-9-4
AUO LCD Cable and Flex Removal Procedure
NOTE:
The numbers shown in square brackets correspond to the numbers listed in
.
1.) Remove the AUO LCD display - refer to the
"AUO LCD Display Removal Procedure" on page 8-55
.
2.) Disconnect the AUO LCD cable
[3]
from the rear of
the LCD panel.
3.) Disconnect and remove the flex board
[4]
by unscrewing the two retaining screws encircled in
.
4.) Disconnect the Inverter Flex Board cable
[2]
from the LCD Backlight Inverter board
[5],
then
disconnect the LCD Back Light Inverter board from the AUO LCD cable
[3]
(arrow at right in
).
5.) Disconnect the Flex Board from the LCD flex
[1]
-
6.) Gently remove the Flex Board and the LCD Back Light Inverter Board.
Figure 8-70 Flex Board and LCD Back Light Inverter Board
Disconnect
here
LCD Backlight
Inverter
Flex Board
2
3
4
5
1
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...