Chapter 8 - Replacement Procedures
8-71
IMPORTANT:
Perform the next steps in the correct order. When securing the ECG Cable Protector, first fasten the rear
securing screw, then the front securing screw.
3.) Return and fasten the
rear
securing screw (
), then the
front
securing screw.
illustrates the ECG Cable Protector secured in position on the SafeLock Cart.
Figure 8-83 Placing the ECG Cable Protector in Position
CAUTION
When placing the ECG Cable Protector in position, take care not to damage the plastic.
Figure 8-84 Fastening the Rear Securing Screw