Chapter 8 - Replacement Procedures
8-119
19.)Press <
Esc>
to return to the
Advanced
menu home screen (
).
20.)Scroll to
Watchdog Settings
and press
<Enter>
to open the
Watchdog Settings
screen.
21.)Set the parameters in the
Watchdog S
ettings as shown above in
.
22.)Press <
Esc>
to return to the
Advanced
menu home screen (
).
23.)Scroll to
Display Control
and press
<Enter>
to open the
Display Control
screen.
24.)Set the parameters in the
Display Control
screen as shown above in
.
Figure 8-122 BIOS Advanced Watchdog Settings
Figure 8-123 BIOS Advanced Display Control Parameters
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...