GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
7-86
Section 7-5 - Monitoring Tests
Section 7-5
Monitoring Tests
7-5-1
Accessing the Monitoring Diagnostic Test Options
•
Access the
Diagnostic
Test
window, as described in the
Accessing the Front End Diagnostic
7-5-1-1
Full System Monitoring Test
NOTE:
For a description of this test, refer to the information in
"Full System Monitoring Test" on page
.
1.) In the
Diagnostic
Test
window, trackball to
Full System Monitoring Test
in the list of
Monitoring
tests available
for selection from the Diagnostic Test Tree (see
) and press
Set
.
The Full System Monitoring Test name is displayed under
Execute;
all system components to be
tested are highlighted light blue in the Schematic View, as shown below:
2.) As indicated in the
Special Instructions
area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the
Special Instructions
area, activate the
Loops
checkbox
and select the required number of
Loops
from the combo box.
4.) Adjacent to
On Fail,
select either
Stop
or
Continue
, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the
Start
button and press
Set
.
The Full System Monitoring Test commences. While the test proceeds, the name of the test currently
in progress (ADC Test) is displayed in the
Execute
field and the current
Loop #
and specific test are
Figure 7-57 Full System Monitoring Test
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...