3-120
Section 3-10 - Connectivity Setup
3-10-6
Query/Retrieve (Q/R) Setup
3-10-6-1
Overview
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM
server for further analysis on the Vivid q N.
3-10-6-2
Query/Retrieve Setup on the Vivid q N
1.) Press
CONFIG
(
F2
) and log on as
Adm
, as described in
"Log On to the System as ADM" on page
.
2.) Select
Connectivity
.
3.) Select
Dataflow
.
4.) Select
Query Retrieve
from the
Name
pull-down menu - see
.
The selected dataflow (Query/Retrieve), is displayed as shown in
.
Figure 3-127 Select Query Retrieve
NAME
PULLDOWN
MENU
QUERY RETRIEVE
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...