Chapter 3 - System Setup
3-65
Section 3-8
Mounting the Vivid q N on the SafeLock Cart (optional)
NOTE:
The Vivid q N portable ultrasound scanner may be mounted on the SafeLock Cart while power
to the system is either ON or OFF. When mounting the unit on the cart, it may be necessary to
adjust the locking mechanism, as described in
3-8-2 "Adjusting the SafeLock Cart Locking
3-8-1
Mounting the Vivid q N on the SafeLock Cart
1.) Working from the front of the SafeLock Cart, place the Vivid q N scanner flat down on the main shelf
(towards the front) then slide it backwards between the sliding guides on either side of the shelf, as
shown below.
NOTICE
Only the SafeLock Cart supplied by GE and/or any medical-grade cart which has been approved as
suitable for use with the system by GE should be used with the Vivid q N portable ultrasound scanner.
NOTICE
The Vivid q N system and the SafeLock Cart should be operated at a distance of more than 15 feet from
any magnetic field.
NOTICE
When scanning with the Vivid q N portable ultrasound scanner mounted on the SafeLock Cart, the
Vivid q N should be connected to the Cart via the USB port.
WARNING
WARNING
Before mounting the Vivid q N portable ultrasound scanner on the SafeLock Cart make
sure there are no foreign objects on the main shelf surface or in front of the locking
tooth.
Figure 3-62 Placing the Vivid q N on the SafeLock Cart Main Shelf
Sliding Guides
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...