Chapter 8 - Replacement Procedures
8-9
The following screen displays:
7.) Select “
Y
” to continue with the formatting process or “
N
” to restart the procedure.
T
he system will start the formatting procedure and reboot automatically.
After rebooting, the system will boot from the software installation CD.
8.) Perform a “Full Clean Install” software procedure corresponding to the BT level of your system.
-
Refer to the
"Software Upgrade Procedure Overview" on page 8-99
Note:
IMPORTANT - if
both
the Hard Disk and Back End are being replaced simultaneously, it is
necessary to re-configure the LCD type.
9.) When done, proceed to perform the following functionality tests:
-
"Software Configuration Checks" on page 4-39
-
"Calibration Tests" on page 7-25
-
"Full System Test" on page 7-10
10.)After you have completed the formatting procedure, store the Formatting Utility Disk at the
customer’s site where it can be accessed for future use.
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...