GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-75
8) Loosen and remove the ground screw fastened to the AC Distribution Assembly - see
9) Loosen and remove the two AC Assembly mounting screws that secure the AC Distribution
Assembly to the mounting bracket on the SafeLock Cart chassis - see
10) Carefully pull the AC Distribution Assembly out towards you and remove it remove from the
SafeLock Cart.
8-4-7-4
AC Distribution Assembly Installation Procedure
1) Return the replacement AC Distribution Assembly to its position on the SafeLock Cart chassis and
secure it on the mounting bracket using the two screws (previously removed - see
2) Fasten the Ground screw to the AC Distribution Assembly -
3) Reconnect the peripheral power cable to the AC Distribution Assembly -
4) Secure all cables in position with new cable ties.
5) Return the AC Assembly cover and fasten the three screws (previously removed).
6) Return the Lower Cover (Rear), as described in the
"Lower Cover (Rear) Installation Procedure" on
7) Return the Upper Cover (Rear), as described in the
"Upper Cover (Rear) Installation Procedure" on
8) Reconnect the AC power cable to the AC Distribution Assembly at the base of the cart (rear) and
turn ON power to the SafeLock Cart.
9) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-88 AC Distribution Assembly and Cable Sockets
Peripheral power
cable
Ground Screw
AC Assembly mounting
screw
AC Assembly mounting
screw
Jumper
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...