3-128
Section 3-10 - Connectivity Setup
3-10-7-3
Configuring the Vivid q N for MPEGVue Export
1) Press the
Config
(or
F2
) button and log on as
Adm
2.) Select
Connectivity
.
3.) Select
Dataflow
.
4) From the drop-down menu, select
Misc Export
.
5) From selected output devices, highlight
evue
and then click the
Properties
button - refer to
.
Figure 3-136 Dataflow - Misc Export