GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-81
8-4-9
Main Cable Harness Replacement Procedure
8-4-9-1
Tools
Appropriate Phillips screwdriver and wire cutters.
8-4-9-2
Preparation
Shut down the Vivid q N ultrasound unit, as described in
Remove the Upper Cover (rear) and Lower Cover (rear) of the SafeLock Cart as described in
Cover (Rear) Removal Procedure" on page 8-66
and
"Lower Cover (Rear) Removal Procedure" on
8-4-9-3
Main Cable Harness Removal Procedure
1) Disconnect the Main Cable Harness from the AC Distribution Assembly and free the cable so that
it does not interfere with the Gas Spring.
2.) Cut any cable ties that secure the cable.
3) Disconnect the P6 cable and USB connector from the USB board, as shown in
.
4) From above the Top Shelf, carefully push the Main Cable Harness Vivid q N power supply cable
through the top shelf aperture and gently pull it through from below.
5) From below the Top Shelf, carefully push the DVD connectors through the front upper cover.
6) Remove the Main Cable Harness from the cart.
8-4-9-4
Main Cable Harness Installation Procedure
1) From the Upper Assembly (rear), thread the Main Cable Harness Vivid q N power supply through
the Top Shelf aperture.
2.) Thread the Main Cable Harness DVD connectors through the aperture of the Upper Assembly
(rear).
3) Wrap the Main Cable Harness once around the Gas Spring and connect the P6 cable and USB
connector to the USB board.
4) Connect the Main Cable Harness to the AC Distribution Assembly.
5) Replace all cable ties to secure the Main Cable Harness.
6) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...